CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS W/SA25 A.CONTROL RESERV.; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX435T
Medical Device Problem Codes Loss of or Failure to Bond (1068); Material Integrity Problem (2978)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fx410t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the patient got worse in (b)(6) 2022.In the x-ray was verified the valve rupture.The patient underwent a revision procedure on (b)(6) 2022.The complaint device has not yet been returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information - a2, a3, d4 (serial #), d9 investigation: visual inspection: the following observations were made during the visual inspection: -loosed spout of the shuntassistant with the connected distal catheter -no visible damage to the valve or spout results: based on our investigation results, we can detect a loss of functionality of the product.The reason for the disconnection of the spout and the gravitational unit/ valve housing is the loose fit.Why the parts showed low contact is subject of further internal investigations in form of a capa.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS W/SA25 A.CONTROL RESERV.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15374803
Report Number3004721439-2022-00274
Device Sequence Number12211518
Product Code JXG
UDI-Device Identifier04041906137335
UDI-Public4041906137335
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2025
Device Model NumberFX435T
Device Catalogue NumberFX435T
Device Lot Number20050026
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer10/28/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient SexMale
-
-