PENUMBRA, INC. INDIGO SYSTEM LIGHTNING 7; QEW Back to Search Results
Catalog Number LITNG7XTORQ130
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was undergoing a thrombectomy procedure to treat a peripheral bypass graft using an indigo system aspiration catheter 7 (cat7), a 7fr non-penumbra sheath, and a guidewire.During the procedure, the physician successfully completed approximately five passes in the target location using the cat7.Upon removal of the cat7 from the patient after completion of the thrombectomy, the physician experienced resistance and broke the cat7 in half and noticed under fluoroscopy that a piece of the cat7 remained in the patient¿s arterial bypass graft.Subsequently, the physician removed the sheath and gained additional access via cutdown technique to open the patient¿s graft and removed the broken distal piece of the cat7.The procedure was completed at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM LIGHTNING 7
Common Device Name
QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15374934
Report Number3005168196-2022-00409
Device Sequence Number17500079
Product Code QEW
UDI-Device Identifier00815948022584
UDI-Public815948022584
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K210323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/07/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2025
Device Catalogue NumberLITNG7XTORQ130
Device Lot NumberH12903
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Initial Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-