Model Number BXA075902A
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 08/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2022, a gore® viabahn® vbx balloon expandable endoprosthesis was intended for use in the common iliac artery for treatment bilateral occlusive disease.However, after the vbx device was successfully delivered, it was noticed on an angiogram, that the internal iliac had been inadvertently covered.The decision was made by the physician to leave the device and monitor the patient.The patient did not experience any adverse consequences.
 
Additional Manufacturer Narrative
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Section h6 updated to reflect completion of investigation.Manufacturing records were reviewed, and the device met all pre-release specifications.The reported unintentional branch vessel obstruction could not be independently confirmed as no items were returned for evaluation.Additional information obtained from the field indicates the physician suspected vessel obstruction was caused by mismeasurement of the target vessel.The gore® viabahn® vbx balloon expandable endoprosthesis instructions for use (ifu), for the appropriate region and time-period.The ifu states the following: ¿special care should be taken to ensure that the appropriate size endoprosthesis, compatible sheath and guidewire are selected prior to introduction.Native vessel dimensions must be accurately measured, not estimated.¿ the ifu also states: ¿once the optimal position is verified fluoroscopically, the endoprosthesis is ready to be deployed.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
nick lafave
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15375185
Report Number2017233-2022-03298
Device Sequence Number5674196
Product Code PRL
UDI-Device Identifier00733132637294
UDI-Public00733132637294
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/17/2024
Device Model NumberBXA075902A
Device Catalogue NumberBXA075902A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexFemale
-
-