MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Failure to Infuse (2340)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "they were able to flush before accessing but once accessed needle was not flushing or aspirating.".
 
Additional Manufacturer Narrative
(b)(4).The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of asfyf085 showed two other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "they were able to flush before accessing but once accessed needle was not flushing or aspirating.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of an occluded infusion set was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 19ga x 1¿ powerloc safety infusion set with y-site.Usage residues were observed throughout the sample and the safety mechanism was engaged.Microscopic inspection of the sample did not reveal any evidence of occlusion.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration with no observed leaks.The effort required to flush the sample was comparable to that required for a similar non-complainant device.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15375226
Report Number3006260740-2022-03493
Device Sequence Number5613593
Product Code LJT
UDI-Device Identifier00801741047893
UDI-Public(01)00801741047893
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/21/2025
Device Model NumberN/A
Device Catalogue Number0671910
Device Lot NumberASFYF085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-