| Model Number |
1258T/92 |
| Medical Device Problem Codes |
Failure to Capture (1081); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2022-21822.It was reported that the patient presented for a follow-up in clinic.Upon interrogation, it was discovered the atrial lead was oversensing noise triggering inappropriate mode switches and the left ventricular lead exhibited a failure to capture caused by dislodgement.The atrial and left ventricular leads were explanted and replaced.The patient was in stable condition before, during, and after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events of lead dislodgement and failure to capture were not confirmed.A partial lead was returned in three pieces for analysis.Electrical and visual analysis were normal with no anomalies found.
|
| |
|