ABBOTT MEDICAL AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number DUCT OCCLUDER
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, a 12-10mm amplatzer duct occluder was selected for procedure.It was noted during procedure that the amplatzer duct occluder exhibited a deformation when being deployed through a 8fr non-abbott delivery sheath.The decision was made to replaced the 12-10mm amplatzer duct occluder with another amplatzer duct occluder of a different unknown size to complete the procedure.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of device deformity was reported.The investigation confirmed the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to the labeling design or manufacturing of the device.Per the instructions for use, the recommended size delivery system for use with a 9-pda-007 amplatzer amplatzer duct occluder is a 7f.
 
Event or Problem Description
It was reported that on 10 august 2022, a 12-10mm amplatzer duct occluder was selected for procedure.It was noted during procedure that the amplatzer duct occluder exhibited a deformation when being deployed through a 8fr non-abbott delivery sheath.The decision was made to replaced the 12-10mm amplatzer duct occluder with another amplatzer duct occluder of a different unknown size to complete the procedure.Subsequent to the previously filed report, additional information was received that: the amplatzer septal occluder exhibited a cobra deformation when being deployed.There was no angulation or kinks noted in the delivery system.The original occluder was never released from the delivery cable and did not interact with any cardiac structures during deployment.The device was replaced with a 10-8mm amplatzer duct occluder (8094671).The patient was reported as stable.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15375631
Report Number2135147-2022-01050
Device Sequence Number5809501
Product Code MAE
UDI-Device Identifier00811806011141
UDI-Public00811806011141
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberDUCT OCCLUDER
Device Catalogue Number9-PDA-007
Device Lot Number6655399
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date11/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8FR MULLINS SHEATH
Patient SexUnknown
-
-