• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DIB00
Medical Device Problem Codes Break (1069); Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age or date of birth, weight, and ethnicity: unknown/ asked but not available.If implanted, give date: not applicable, as there was no patient involvement.If explanted, give date: not applicable, as there was no patient involvement the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that preloaded intraocular lens (iol) had distorted haptic and the trailing haptic was not tucked and sticking out which could not be adjusted.The haptic was facing the wrong way which could not be turned/rotated without scratching/damaging the lens.The account believed it was because the haptic was damaged.There was no other issue with the lens and the procedure was completed using another lens.There was no patient contact with the original lens.There was no user error involved.No other information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes section d9: returned to manufacturer on: oct 7, 2022 section h3: device evaluated by manufacturer: yes device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint handpiece was received with the complaint lens stuck inside of the cartridge and with the lens overridden by the plunger rod.The trailing haptic could also be observed to be unfolded.Furthermore, the cartridge tip could also be observed to be bent, in a way consistent with a handpiece that was dropped.The handpiece was disassembled and the assembly was inspected, no issues that could contribute to or cause the observed issues were identified.The complaint lens was removed from the cartridge and cleaned, revealing creases from being folded inside of the cartridge.No further issues were observed.Conclusion: the reported complaint issue of trailing haptic not folded was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other reported complaint issue haptic damaged was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15375695
Report Number3012236936-2022-02332
Device Sequence Number5661408
Product Code HQL
UDI-Device Identifier05050474731691
UDI-Public(01)05050474731691(17)240415
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2024
Device Model NumberDIB00
Device Catalogue NumberDIB00U0170
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer10/18/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-