It was reported that the bd stopcock¿ leaked from the q-syte connection while infusing campto and nipro.The following information was provided by the initial reporter, translated from japanese: "during infusion of the anticancer agent campto, along with levofolinate (nipro) through the side route, leakage from the stopcock was observed; this event was noticed when the infusion was supposed to be completed in about 30 min.According to the nurse¿s report, the q-syte seemed to be connected diagonally.Reconnection was attempted but the fluid leakage lased and the iv line was replaced.".
|
Bd was unable to perform a thorough investigation as no sample, lot, or batch number were provided.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.Examination of the product involved may provide clarification as to the cause for the reported failure.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|