• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. SLEEK OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. SLEEK OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number 426-1512X
Medical Device Problem Codes Material Rupture (1546); Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure through popliteal artery, the balloon allegedly ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 10/2024).Device pending return.
 
Additional Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the physical sample was returned for evaluation.An attempt was made to inflate the balloon with water but this was unsuccessful.The catheter outer exhibited evidence of stretching commencing at the proximal cone for a length of 140mm.It is likely that the stretched section prevented water reaching the balloon during the inflation attempt.The visual inspection of the balloon did not reveal any evidence of a rupture.The patency of the device using a 0.014" guidewire was also unsuccessful due to the stretched section.The stretched section is likely user related and unlikely manufacturing related as a 100% visual inspection for defects and also guidewire check is performed during catheter production.The result of the investigation is inconclusive for the reported balloon rupture issue.The definitive root cause for the reported balloon rupture issue could not be determined based upon the available information received from the field communications and device evaluation.Labeling review: the instructions for use for the sleek otw product was reviewed and contains the following information relevant to the reported event: balloon characteristics: please check the package label for the rated burst pressure (rbp).It is important that the balloon not be inflated beyond the rated burst pressure.Pressures in excess of rated burst pressure may cause the balloon to burst.Warnings: do not exceed the rated burst pressure.A syringe with pressure gauge is recommended to monitor pressure.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Use a 20 ml or larger syringe for inflation.Do not advance the guidewire, balloon dilation catheter, or any component if resistance is met, without first determining the cause and taking remedial action.This catheter is not recommended for pressure measurement or fluid injection.Precautions: carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.If resistance is felt upon removal, then the balloon, guidewire and the sheath/guide catheter should be removed together as a unit, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and sheath/guide catheter as a unit and withdrawing both together, using a gently twisting motion combined with traction.Before removing catheter from sheath/guide catheter it is very important that the balloon is completely deflated.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.Directions for use: inspection and preparation: prepare a mixture of contrast medium and normal saline as per normal procedure.(recommended 25%/75%).Attach a stopcock and a 20 ml syringe half filled with the contrast solution to the balloon port.Point the syringe nozzle downward and aspirate until all air is removed from the balloon.Turn the stopcock off and maintain the vacuum in the balloon.Purge the catheter through lumen thoroughly.Reinserting the balloon into the balloon sleeve may damage the balloon or catheter.Procedure: insertion and inflation procedure: note: a 0.014¿ (0.356 mm) guidewire must be inserted in the sleek otw catheter across the balloon during any inflation of the balloon.Make sure that the balloon sleeve has been removed from the catheter balloon.Enter the vessel percutaneously using the standard seldinger technique over the appropriate guidewire for the size catheter being used.Advance the catheter across the lesion with fluoroscopic guidance using accepted percutaneous transluminal angioplasty technique and inflate the balloon to the appropriate pressure.Note: do not inflate the balloon or advance the catheter unless the guidewire is in place.H10: d4 (expiry date: 10/2024), g3, h6 (device).H11: h6 (method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during an angioplasty procedure through popliteal artery, the balloon allegedly ruptured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLEEK OTW PTA CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15376451
Report Number9616666-2022-00191
Device Sequence Number14868173
Product Code DQY
UDI-Device Identifier00801741133558
UDI-Public(01)00801741133558
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K102035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number426-1512X
Device Catalogue Number426-1512X
Device Lot NumberCMFY0050
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-