| Model Number |
40458 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that guidewire fractured at the tip inside the patient.This v-14 control wire was selected for dilatation of dialysis shunt in the arcus palmaris superficialis.The vessel was 80% stenosed with severe calcification and moderate tortuosity.During the placement of this guidewire through the stenosis, when the physician pulled back, the tip broke.A snare was used to try to remove the guidewire, however it was only partially able to be explanted and a portion remains implanted.No patient complications were reported, and the procedure was not able to be completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by manufacturer: this product was not returned for analysis, as the device was disposed of by the physician.An x-ray was submitted for review that confirmed the distal end of the guidewire was detached/separated from the rest of the device.
|
| |
|