| Model Number |
1219-28-150 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Pain (1994); Scar Tissue (2060); Physical Asymmetry (4573)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Surgeon reported to bf arm: "on 08-aug the patient had sudden groin pain on the right side, since 10-aug there is a squeaking noise in the area of the right hip joint.The patient presented herself yesterday in the doctor's office; the indication for a hip tep replacement because a decentered head in the cup was given.(b)(6) 2017 implantation of a cementless total endoprosthesis right hip (corail stem size 10, head size 28/+5, pinnacle cup size 50/28 mm).(b)(6) 2022.Explantation of the broken inlay.The inlay was also luxated." doi: (b)(6) 2017.Dor: (b)(6) 2022.Right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
There was a leg shortening of approx.1 cm and the scar conditions were unattractive.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was received for analysis.Based in the visual examination of the part, it was observed that a portion of the liner's rim was fractured.Location of the fracture is consistent with excessive edge loading of the femoral head and patient's body weight since machining lines are yet visible opposite to the fracture zone.Interlock feature edge of the liner denoted three ard tabs sheared off from the liner which is also evidence of the liner having disassociated from the cup.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : product description: pinn mar +4 10d 28idx50o product code: 121928150 lot number: h24095 please provide: 1) quantity manufactured: 40 2) date of manufacture: 12/20/2016 3) any anomalies or deviations identified in dhr: none.No anomalies or non conformances were identified.4) expiry date: 11/30/2021 5) ifu reference: 090200701 - ifu tot hip pros rev p.Device history review : a manufacturing record evaluation was performed for the finished device 121928150/h24095) product and lot numbers, and no non-conformances were identified.Corrected: h3.
|
| |
|
Search Alerts/Recalls
|