• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, AARDVARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, AARDVARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279251101
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device broke during the procedure.The broken piece was retrieved.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: a piece of the white part at the tip of the wand had broken off and was floating in the joint space.The probable root cause/s could be excessive force used when inserting removing the probe, a probe inserted into an obstructed passageway or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the device broke during the procedure.The broken piece was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES, AARDVARK
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15379060
Report Number0002936485-2022-00505
Device Sequence Number5707856
Product Code GEI
UDI-Device Identifier07613327056884
UDI-Public07613327056884
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0279251101
Device Catalogue Number0279251101
Device Lot Number21340AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-