| Model Number |
72202067 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an unknown procedure, the twinfix anchor was introduced without problems, but when the sutures were passed through the tissues and the knots were tied, they broke without making a major effort.This happened with the two sutures, so the anchor could not be used.Surgery was resumed, after a non-significant surgical delay, with a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A review of the drawings found that that material certification of analysis is require for both sutures.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The twinfix suture anchor is implantable, biocompatibility is not an issue.If retained micro-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|