| Catalog Number |
EN2007030/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a ureteric stent removal procedure, the distal portion of a vascular snare device detached within the patient during the removal processes.The physician successfully externalized the foreign body with another vascular snare device liberating the vessel.The procedure was successfully completed for this patient.No additional consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for investigation.The device was examined visually and microscopically.The complaint is confirmed.The root cause could not be determined however, it is likely that significant force was applied to the device during clinical use.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
| |
|