• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ZEBRA; STYLET FOR CATHETER, GASTRO-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ZEBRA; STYLET FOR CATHETER, GASTRO-UROLOGY Back to Search Results
Model Number M0066701091
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a zebra guidewire was used during holmium laser lithotripsy procedure on (b)(6) 2022.During preparation, one guidewire in a box of five was not sealed and had an opening.The sterile barrier was compromised however there was no visible damage noted on the outer packaging.The procedure was not completed due to this event and device was out of stock.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Initial reporter state:(b)(6).(b)(4).
 
Event or Problem Description
T was reported to boston scientific corporation that a zebra guidewire was used during holmium laser lithotripsy procedure on (b)(6) 2022.During preparation, one guidewire in a box of five was not sealed and had an opening.The sterile barrier was compromised however there was no visible damage noted on the outer packaging.The procedure was not completed due to this event and device was out of stock.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block e1 (initial reporter state): (b)(6).Block h6 (device): device code a020503 captures the reportable event of unsealed device packaging.Block h10: a zebra guidewire was returned.The pouch returned open with evidence that it was manipulated, it was found that pouch head seal rubber residues for both, the chevron part (supplier seal) and the tyvek seal part (bsc seal).In addition, in the analysis, rubber residues of a label were found in the front section, and the pouch came closed with a staple.Therefore, the reported event was not confirmed.Based on a review of all available information, the cause of the reported event could not be determined.The product record review results and external support form meets all manufacturing specifications required and passed all the controls and inspections; no abnormalities were reported during the assembly process.Since the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event, cause not established is selected as the most probable cause for the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEBRA
Common Device Name
STYLET FOR CATHETER, GASTRO-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15380261
Report Number3005099803-2022-05109
Device Sequence Number5661559
Product Code EZB
UDI-Device Identifier08714729191223
UDI-Public08714729191223
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2024
Device Model NumberM0066701091
Device Catalogue Number670-109
Device Lot Number0027057033
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age13 YR
Patient SexMale
Patient Weight45 KG
-
-