• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SYNTEL SILICONE OVER-THE-WIRE EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SYNTEL SILICONE OVER-THE-WIRE EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Model Number A4E01
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received the complaint device for evaluation and have confirmed the reported incident.The device was inspected and the balloon was confirmed to be ruptured.The product was in use at the time of the failure.Inspection of the device could not determine a precise root cause for the failure.It is possible that the balloon came in contact with a calcification during the procedure.There was no patient injury resulting from the event.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.
 
Event or Problem Description
The syntel silicone over-the-wire embolectomy catheter was used in a thrombectomy of a dialysis patient.After preparing the product, the catheter tip was inserted into the vein.When the balloon was inflated and the thrombus removal was started, the balloon ruptured.The procedure was completed using another catheter without any problems.The patient is fine.No health damage happened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
jeffrey oddy
63 second ave
burlington, MA 01803
7814251682
MDR Report Key15381440
Report Number1220948-2022-00093
Device Sequence Number5648505
Product Code DXE
UDI-Device Identifier00840663109869
UDI-Public00840663109869
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA4E01
Device Catalogue NumberA4E01
Device Lot NumberPYT1066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Initial Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-