• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 07797826160
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
K.N 12/22/2022 the reporter complained of discrepant inr results with coaguchek xs meter serial number (b)(4) compared to an unknown laboratory method.The result from the meter was 2.3 inr.The result from the laboratory was 1.8 inr.On (b)(6) 2022 at approximately 5:04 pm, the meter result was 2.6 inr.The result from the laboratory was 2.0 inr.On (b)(6) 2022 at 4:15 pm, the result from the meter was 3.1 inr.The result from the laboratory at the same time was 2.3 inr.The patient's therapeutic range was unknown.
 
Additional Manufacturer Narrative
The test strips were requested for investigation.The reporter¿s strips were provided for investigation where they were tested using a retention meter and retention controls.Testing results (qc range = 2.4 - 3.6 inr): qc 1: 2.8 inr, qc 2: 2.8 inr, qc 3: 2.7 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.Customer is wiping away the first drop of blood and applying the second drop of blood.They stated they often squeeze the finger a lot to get the second blood drop.Per product labeling, "if needed, immediately after lancing, gently massage the finger from its base until a drop of blood is formed.Do not press or squeeze the finger." on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Product labeling states: "the coaguchek system uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas deviations can occur when compared to methods using other thromboplastins.However, those deviations between thromboplastins of different origin (e.G., rabbit based) are not specific to coaguchek products.Similar differences can be observed when a human recombinant thromboplastin based laboratory method is compared to other laboratory methods." the investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15382266
Report Number1823260-2022-02758
Device Sequence Number5707403
Product Code GJS
UDI-Device Identifier04015630946471
UDI-Public04015630946471
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number07797826160
Device Lot Number58943516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Initial Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN; HYDROXYZINE; IRON; MIRTAZAPINE; WARFARIN
Patient Age50 YR
Patient SexFemale
Patient Weight60 KG
-
-