| Catalog Number |
03P88-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Cardiogenic Shock (2262)
|
| Date of Event |
08/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 29-aug-"2021", abbott point of care was contacted by a customer regarding i-stat cg8+ cartridges that yielded a suspected discrepant ionized calcium result of >250 mmol/l on a 60 year old female patient with cardiogenic shock.Return product is available for investigation.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.Product lot information was not presented, but requested.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4) the investigation was completed on 22-sep-2022.A review of the device history record confirmed the lot passed finished goods release criteria.The results from retained and returned cartridge testing of cg8+ cartridge lot w22170 met the acceptance criteria found in q04.01.003 rev.Ak product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
|
| |
|