| Catalog Number |
03P76-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022 abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected descrepant hemglobin results on a 28 year old female patient.There was no additional patient information available at the time of this report.Return product is available for investigation.Method date collected tested hgb, i-stat 30-aug-2022 13:40 13:48 11.6 g/dl, sysmex 30-aug-2022 ni ni 9.8 g/dl, i-stat 30-aug-2022 ni 13:52 13.9 g/dl, sysmex 30-aug-2022 ni ni 13.1 g/dl.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 13-sep-2022.A review of the device history record confirmed the lot passed finished goods release criteria.Retained and returned cartridge testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|