• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES DIC; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES DIC; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 513070
Medical Device Problem Codes Inflation Problem (1310); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trach does not inflate, when put in water, you can see there is a hole in it.Not used on a patient.
 
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other, other text: one device was received for investigation.Visual inspection did not reveal any defects.Leakage was observed immediately during the leak test.Further investigation using magnification revealed a slight tear in the cuff material; the complaint was confirmed.While the leakage of the cuff was confirmed it could not be determined with any certainty at what point this defect occurred.Root cause could not be determined.One year complaint history review since september 2021 revealed no trends for this product and problem category.Complaint information will continue to be monitored for any new information or adverse trends and further actions will be taken accordingly.There were no non-conforming reports (ncrs) initiated for the lot or anomalies identified in the device history record review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES DIC
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15383272
Report Number3012307300-2022-18218
Device Sequence Number17500161
Product Code BTR
UDI-Device Identifier10351688023544
UDI-Public10351688023544
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K934465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number513070
Device Catalogue Number513070
Device Lot Number4168591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-