BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00517450
Medical Device Problem Codes Difficult to Remove (1528); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that an agile esophageal stent was to be implanted in the esophagus to treat an approximately 4-5 cm esophageal tumor during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, the agile esophageal stent was successfully deployed under fluoroscopic guide.Upon removal of the agile delivery system, the stent suture loop got caught on the delivery system and the agile esophageal stent moved out from the target location.The agile esophageal stent was removed together with the delivery system through the patient's mouth.The procedure was completed with another agile esophageal stent.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILE ESOPHAGEAL
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15384976
Report Number3005099803-2022-05116
Device Sequence Number5613886
Product Code ESW
UDI-Device Identifier08714729973072
UDI-Public08714729973072
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K180144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/13/2024
Device Model NumberM00517450
Device Catalogue Number1745
Device Lot Number0029225826
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2022
Initial Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-