C.R. BARD, INC. (BASD) -3006260740 POWERPICC CATHETER 4F FULL TRAY (SINGLE-LUMEN); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer " during removal of a picc line from a patient the line was inspected and noted to have tear in the catheter wall around the 30 cm mark.The tear was approximately 1 cm in length and the end of the catheter was intact.It did not appear that any of the catheter was retained within the patient, no harm to the patient was identified".
 
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of refw4256 showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged catheter was confirmed.The product returned for evaluation was one 4 fr single lumen powerpicc.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated, and a split was observed in the catheter at the 30 cm depth marker.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: a relatively large split with flared edges.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).An occlusion was observed distal to the burst site, and this may have contributed to the failure.Forceful flushing against an occlusion can cause this type of failure.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.
 
Event or Problem Description
It was reported by the customer " during removal of a picc line from a patient the line was inspected and noted to have tear in the catheter wall around the 30 cm mark.The tear was approximately 1 cm in length and the end of the catheter was intact.It did not appear that any of the catheter was retained within the patient, no harm to the patient was identified.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by the customer " during removal of a picc line from a patient the line was inspected and noted to have tear in the catheter wall around the 30 cm mark.The tear was approximately 1 cm in length and the end of the catheter was intact.It did not appear that any of the catheter was retained within the patient, no harm to the patient was identified" additional information received 7/21/2023: customer reported to bd medical affairs that this event involved a teleflex wire; the customer did not use the bd stylet from the kit to insert the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC CATHETER 4F FULL TRAY (SINGLE-LUMEN)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15385087
Report Number3006260740-2022-03518
Device Sequence Number20282849
Product Code LJS
UDI-Device Identifier00801741027574
UDI-Public(01)00801741027574
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K070996
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberN/A
Device Catalogue Number3174108
Device Lot NumberREFW4256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer10/14/2022
07/21/2023
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date10/19/2022
07/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-