MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.5/15; COVERED CORONARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.5/15; COVERED CORONARY STENT Back to Search Results
Model Number 434889
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected pk papyrus stent systems were not returned to biotronik and could therefore not be subjected to a technical investigation.Clinical information (e.G.Cathlab report) or images of the case such as angiographies could also not be obtained.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted review no manufacturing related root cause could be identified.The 6f guideliner extension catheter used in the intervention is compatible with the affected pk papyrus.However, when using an extension catheter, it should be considered that inside the guiding catheter more edges are present and the risk to be caught on one of the edges is increased.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Event or Problem Description
It was attempted to put multiple pk papyrus stents through a guide liner that was not large enough to accommodate the stents.Three 3.5/15 pk papyrus came off, two in the guide and one in the sheath.They were all removed without any patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 3.5/15
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15385235
Report Number1028232-2022-04588
Device Sequence Number5429714
Product Code NIV
UDI-Device Identifier07640130447172
UDI-Public(01)07640130447172(17)2308
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number434889
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07213403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age85 YR
Patient SexFemale
-
-