Model Number B2060-150
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat the right peritoneal trunk artery with moderate calcification.The 6x150 armada 35 percutaneous transluminal angioplasty (pta) catheter was advanced to the lesion and inflation to nominal pressure was attempted; however, the balloon would not inflate at all.The device was removed and a new, same size armada 35 was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.Returned device analysis identified that the balloon inflated but was unable to maintain pressure due to fluid leaking at the proximal end of the sidearm.No additional information was provided.
 
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported balloon inflation issue was confirmed.Additionally a leak in the system was also noted.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the noted leak in the system or reported inflation issue.However, factors that can contribute to a leak in the system include, but are not limited to, manufacturing, interactions with accessory devices, catheter damage, or insufficient preparation prior to use.The product risk assessment identifies this as a foreseeable event; as such, design and manufacturing controls and mitigations have been established for potential causes.In this case, it is likely that the noted leak resulted in the reported inflation issue.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15385542
Report Number2024168-2022-09525
Device Sequence Number12165902
Product Code LIT
UDI-Device Identifier08717648155314
UDI-Public08717648155314
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberB2060-150
Device Catalogue NumberB2060-150
Device Lot Number10504G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR GUIDE CATHETER
Patient SexUnknown
-
-