MAUDE Adverse Event Report: BIOMET 3I 3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number BOST611
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Nerve Damage (1979); Numbness (2415); Paresthesia (4421)
Date of Event 05/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed due to the patient noting numbness and nerve injury.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).Weight unknown / not provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
One 3i t3® non-platform switched tapered implant 6 x 11.5mm (bost611) was returned for investigation.Visual evaluation of the as returned product identified signs of use but no apparent signs of malfunction.The device could not be functionally tested for the reported failure/event (medical: other ¿ numbness/nerve injury/paresthesia).However, dimensional analysis has confirmed that the product was within specification when it left zimvie.Pre-existing condition noted in the per was high bone density ¿type i.The reported device had been placed on tooth #19 for approximately 1 month.X-ray & picture evaluation: none provided.Appropriate documentation was reviewed.Dhr review for the lot (2020120541) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op#0210) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (2020120541) for similar events and no other complaint was identified.November post market trending was reviewed and there were no actionable events or corrective actions for the reported event or product.Therefore, based on the available information, device malfunction did not occur, and the reported event could not be verified as the exact details of event and patient anatomical conditions were unknown/unverifiable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15385703
Report Number0001038806-2022-01393
Device Sequence Number17594824
Product Code DZE
UDI-Device Identifier00844868029717
UDI-Public(01)00844868029717(17)260114(10)2020120541
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/14/2026
Device Model NumberBOST611
Device Catalogue NumberBOST611
Device Lot Number2020120541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-