• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RELIGN CORPORATION SHAVER HANDPIECE ATCLV; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RELIGN CORPORATION SHAVER HANDPIECE ATCLV; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number N/A
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Smoking (1585); Suction Problem (2170); Sparking (2595)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Procode: hrx, gei.The customer has indicated that the product will be returned to zimmer biomet and the investigation is in process.Once the investigation has been completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported the device wasn't working correctly and the suction was off, so they pulled it out of the patient and hit the shaver mode to check it out.The device sparked while oscillating.The handpiece was switched out and everything worked great after that and patient was not harmed.It was reported that no further information is available.No adverse events were reported as a result of this malfunction.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information available.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: the device was observed to have rust and damage on the motor shaft bearing causing resistance when the shaft was turned.The corrosion is evidence of fluid ingress into the motor.The handpiece was dispositioned for scrap.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAVER HANDPIECE ATCLV
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
RELIGN CORPORATION
655 campbell technology pkwy
suite 275
campbell CA 95008
Manufacturer (Section G)
RELIGN CORPORATION
655 campbell technology pkwy
suite 275
campbell CA 95008
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15385948
Report Number3012790575-2022-00004
Device Sequence Number14902198
Product Code GEI
UDI-Device Identifier00859273007236
UDI-Public(01)00859273007236(11)210830
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K162770
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberR10023
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/30/2022
01/17/2023
Initial Report FDA Received Date09/08/2022
Supplement Report FDA Received Date10/07/2022
01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-