| Model Number |
VTICM513.2 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm vticm513.2 with a, -11.50/2.0/071 (sphere/cylinder/axis) into the patient's eye.It is reported, "lens to big they will remove and implant a smaller size.".
|
| |
|
Additional Manufacturer Narrative
|
|
Patient information unk.Work order search found no simliar complaints.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5: lens was exchanged on (b)(6) 2022 with a shorter length lens and problem was resolved.Reportedly, "looks great." cause was reported as patient-related factor and device failed to perform as intended.Patient did not conform to sizing nomogram.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned dry in a ziploc.Visual inspection found residue on a haptic torn lens.Lens model was not a vticm5.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|