MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 0.14 GUIDEWIRES; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 0.14 GUIDEWIRES; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 103-0606-200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that the avigo guidewire broke in the patient's body, and a thrombectomy stent was used to remove it.The guidewire was replaced, and the surgery was successfully completed. the patient did not experience any injury.  the patient was undergoing treatment for acute ischemic stroke.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that there had been no friction/difficulty during delivery, and no issues that resulted in the damage that was found.There was no known reason for the break.The patient's vessel tortuosity was normal.The devices were prepared as indicated in the instructions for use (ifu).
 
Additional Manufacturer Narrative
H3: the avigo guidewire was returned separated into two segments.The avigo guidewire was found bent at ~6.1cm from broken end.In addition, the avigo guidewire was found to be broken at ~5.0cm from distal tip.No other anomalies were observed.The avigo guidewire total length was measured to be within specification after measured both segments.It appears all segments of the avigo guidewire were returned.There does not appear to be any missing segments.The guidewire broken end will be sent out to element labs for sem (scanning electron microscopy) analysis.Based on the device analysis and reported information, the customer¿s report of ¿guidewire separation/break¿ was confirmed as the avigo guidewire was returned broken.However, the root cause could not be determined.Per the sem analysis, the wire failed via ductile overload.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
0.14 GUIDEWIRES
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15388273
Report Number2029214-2022-01525
Device Sequence Number14905381
Product Code DQX
UDI-Device Identifier00847536004764
UDI-Public00847536004764
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2023
Device Model Number103-0606-200
Device Catalogue Number103-0606-200
Device Lot NumberB191499
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer09/16/2022
11/12/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/04/2022
11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight62 KG
-
-