We checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, we confirmed that the segment fluid damage, the light guide fiber bundle (lcb) broken, the air nozzle clogged, the water nozzle clogged, the control body corroded, the insertion flexible tube (ift) leak, the angulation-down angulation decrease, the angulation-left angulation decrease, the angulation-right angulation decrease, the angulation-up angulation decrease, and the u/d and r/l knobs white marking faded/missing; however, they are not the main cause, and/or irrelevant to the alleged complaint.In terms of gluing missing, the possibility of dropping into human body could not be denied.Moreover, based on the technical report""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated to submit mdr.
|