• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1210 SYNCHRONY ST
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user reported that recently the implant housing has moved down towards the pinna.A revision surgery for an implant bed is considered.
 
Event or Problem Description
In (b)(6) 2022 the user had some swelling around the implant site, which was resolved with antibiotic treatment.The user also reported that the implant housing has moved down towards the pinna and when he eats, laughs or touches his pinna he hears a disturbing whistling sound.The user was re-implanted.
 
Additional Manufacturer Narrative
Conclusion: according to the information received from the field the implant housing migrated from its intended position.Further the recipient suffered from a swelling above the implant site, which was resolved with antibiotics.During explantation mastoidectomy cavity was found totally closed with new grown bone.In addition device investigation revealed damage to the active electrode caused by device minute mobility.Reported events might have contributed to the device migration.The reported housing migration likely led to movement of the electrode causing fatigue wire breakages.Other mechanical damages found are attributable to the removal surgery.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key15388783
Report Number9710014-2022-00662
Device Sequence Number5810426
Product Code MCM
UDI-Device Identifier09008737331832
UDI-Public(01)09008737331832
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/17/2021
Device Model NumberMI1210 SYNCHRONY ST
Device Catalogue Number33334
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer09/06/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date03/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
Patient SexMale
-
-