• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJO (SUZHOU) CO., LTD. NIMBUS 4; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJO (SUZHOU) CO., LTD. NIMBUS 4; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE Back to Search Results
Model Number 650001DAR
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code Fall (1848)
Date of Event 08/28/2022
Type of Reportable Event Death
Event or Problem Description
Arjo became aware of an event involving a nimbus 4 mattress.A paraplegic patient fell out of the mattress and was found on the floor.He was lifted back into the bed using the hoist.The patient's condition had progressively deteriorated and the patient passed away.
 
Event or Problem Description
Arjo became aware of an event involving a nimbus 4 mattress.According to the report forwarded by the police to arjo, a paraplegic patient fell out of the mattress and was found on the floor.He was lifted back into the bed using the hoist.The patient's condition had progressively deteriorated and the patient passed away.The nimbus mattresses consist of a top cover which covers the mattress support surface consisting of the sheet with air-filled cells under which is located the automatt sensor pad (mattress base).Based on the photographic evidence provided, the mattress was over-inflated what contributed to the patient fall from the device.
 
Additional Manufacturer Narrative
Additional information was provided as the involved mattress was returned to arjo on 12 oct 2022 what allowed to establish model number and serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIMBUS 4
Common Device Name
MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE
Manufacturer (Section D)
ARJO (SUZHOU) CO., LTD.
no. 158 fangzhou road, sip
suzhou, jiangsu 21502 4
CH  215024
MDR Report Key15389653
Report Number1419652-2022-00040
Device Sequence Number14902233
Product Code FNM
UDI-Device Identifier05055982782234
UDI-Public(01)05055982782234(11)160304
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
Report Date (Section F) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number650001DAR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/02/2022
Device Age6 YR
Event Location Nursing Home
Date Report to Manufacturer11/03/2022
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date11/04/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexMale
Date Report Sent to FDA11/03/2022
-
-