| Model Number |
650001DAR |
| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
08/28/2022
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
Arjo became aware of an event involving a nimbus 4 mattress.A paraplegic patient fell out of the mattress and was found on the floor.He was lifted back into the bed using the hoist.The patient's condition had progressively deteriorated and the patient passed away.
|
| |
|
Event or Problem Description
|
|
Arjo became aware of an event involving a nimbus 4 mattress.According to the report forwarded by the police to arjo, a paraplegic patient fell out of the mattress and was found on the floor.He was lifted back into the bed using the hoist.The patient's condition had progressively deteriorated and the patient passed away.The nimbus mattresses consist of a top cover which covers the mattress support surface consisting of the sheet with air-filled cells under which is located the automatt sensor pad (mattress base).Based on the photographic evidence provided, the mattress was over-inflated what contributed to the patient fall from the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided as the involved mattress was returned to arjo on 12 oct 2022 what allowed to establish model number and serial number.
|
| |
|