MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. REDIGUARD IAB: 7FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. REDIGUARD IAB: 7FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN039783
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Mdr report key/report number 3600840000-2022-8070.
 
Event or Problem Description
Complaint from medsun report: it was reported "early 60's female with history of diabetes, hypertensoin and acute chest pain.Procedure: cardiac cath, percutaneous intervention, stent placement and balloon angioplasty.While doing the procedure, the balloon would not inflate.Device removed and another one used without difficulty.No known harm to patient".
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of iab would not inflate completely is not able to be confirmed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Additional information received stated that the "patient was discharged 02 aug 2022 without known issues from the device".
 
Event or Problem Description
Complaint from medsun report: it was reported "early 60's female with history of diabetes, hypertensoin and acute chest pain.Procedure: cardiac cath, percutaneous intervention, stent placement and balloon angioplasty.While doing the procedure, the balloon would not inflate.Device removed and another one used without difficulty.No known harm to patient".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDIGUARD IAB: 7FR 30CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15390464
Report Number3010532612-2022-00360
Device Sequence Number17546219
Product Code DSP
UDI-Device Identifier10801902145649
UDI-Public10801902145649
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K981660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/31/2023
Device Model NumberIPN039783
Device Catalogue NumberIAB-S730C
Device Lot Number18F21H0001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient Age60 YR
Patient SexFemale
Patient Weight68 KG
-
-