• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 3.5, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 3.5, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT TIB IMP CRC INSERT SZ 3.5, 9MM
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Failure of Implant (1924); Pain (1994); Synovitis (2094)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by mfr: pending investigation.Concomitant medical products: 02-020-13-0235, (b)(4)- truliant cr cem fem cr cem left sz 3.5.02-022-45-3535, (b)(4)- truliant tib fit tray cem sz 3.5f / 3.5t.200-02-38, (b)(4)- three peg patella 38mm.521-78-23, (b)(4)- threaded pin size 2.3 collared 2pk.A10012, (b)(4) - gps implant kit v2.
 
Event or Problem Description
As reported, the patient had an initial tka on (b)(6) 2018.The patient presented to the clinic with pain.Mri showed synovitis.Revision surgery was performed on (b)(6) 2022.The surgeon released the scar tissue around the anterior knee.He noted only mild surface changes to the insert.A new 9mm insert was implanted with good rom and stability.The patient was last known to be in stable condition following the event.Explants are not available without patient release.No further information available.
 
Additional Manufacturer Narrative
H3: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the pain and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.H7: z-0023-2022.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c, d, g.The reason for the revision reported cannot be confirmed from the information provided but may be the result of the reported pain or due to inclusion of the polyethylene in the packaging recall.Prosthesis wear of the tibial insert could not be confirmed from the provided image and radiograph.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP CRC INSERT SZ 3.5, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15391230
Report Number1038671-2022-01070
Device Sequence Number6124222
Product Code JWH
UDI-Device Identifier10885862307699
UDI-Public10885862307699
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/29/2026
Device Model NumberTRULIANT TIB IMP CRC INSERT SZ 3.5, 9MM
Device Catalogue Number02-022-51-3509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer10/11/2022
01/13/2025
07/07/2025
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/29/2022
01/24/2025
07/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-