U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-CT SOMATOM DEFINITION AS; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-CT SOMATOM DEFINITION AS; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 08098027
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to siemens that an adverse event occurred while operating the somatom definition as ct system.On (b)(6), 2022, a 1 year old infant patient underwent examination.The head spiral scan was aborted because someone accidentally opened the door to the gantry room and the door switch interrupted the examination.According to available information, the head spiral scan stopped after 2.23 seconds from the planned 5.46 seconds and the examined part could not be reconstructed.A rescan was needed to complete the examination.There was no delay in diagnosis as the customer started a second spiral scan immediately after the first scan to complete the examination.The health status of the infant is well.No negative health consequences were reported in association with the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.Based on the information available, no general product design issue could be identified, and this issue is rated as a handling issue.The system works as specified.Therefore, this report has been submitted with an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION AS
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key15392227
Report Number3004977335-2022-42540
Device Sequence Number17758246
Product Code JAK
UDI-Device Identifier04056869003665
UDI-Public04056869003665
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number08098027
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age1 YR
Patient SexUnknown
-
-