SIEMENS HEALTHCARE GMBH-CT SOMATOM DEFINITION FLASH; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 10430603
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to siemens that a malfunction occurred prior to operating the somatom definition flash ct system.On (b)(6) 2002, two children (15 months and 7 years) were planned for examination.A system malfunction occurred and the examination could not be performed.The siemens service engineer checked the system onsite and found two defective fuses on the pdr component (power distribution rotating) and replaced them.Further checks were performed, and a defective tube was also identified.The tube was replaced on (b)(6) 2022, and the system was brought back to clinical operation.According to information received, both patients were rescanned and are in good health condition.No negative health consequences were reported for the children associated with the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The material consumption in relation to the installed base is monitored by our capa process.The consumption for the affected component on this specific system is normal and within defined thresholds.No further investigation within the complaint process is necessary.Based on the information available, no general product design issue could be identified, and no further measures are required.This issue is rated as a service case.Therefore, this report has been submitted with an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION FLASH
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key15392235
Report Number3004977335-2022-42881
Device Sequence Number6463780
Product Code JAK
UDI-Device Identifier04056869006963
UDI-Public04056869006963
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10430603
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age7 YR
Patient SexUnknown
-
-