| Model Number |
1011-0139-05 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated device failed to discharge using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation; instead, a zoll representative went onsite to evaluate the internal handles.The customer's report was duplicated and attributed to a faulty shock button on the internal handles.The internal handles will be scrapped and replaced.Analysis for reports of this type has not identified an increase in trend.
|
| |
|