• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN® LOCATOR® ABUTMENT 3.4MM(D) X 5MM(H); DENTAL SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN® LOCATOR® ABUTMENT 3.4MM(D) X 5MM(H); DENTAL SCREW Back to Search Results
Model Number IMLOA005
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number: (b)(4).Age at time of the event: unknown / not provided.Patient weight: unknown / not provided.Additional device information: unknown / not provided.Device manufacturer date: unknown / not provided.
 
Event or Problem Description
The doctor reports that the patient goes to his medical office with the prosthesis removed and there the doctor shows the locator breakage in the part of the body.The affected tooth position is #36.The dental procedure was not completed with another item.The patient did not suffer any type of negative impact on his health as a consequence of the event.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The complaint report was received, reviewed, and evaluated by the manufacturer's complaint process.Applicable manufacturer¿s internal lot history records were reviewed to evaluate whether any internal deviations, non-conformances, or problems occurred during the manufacture of the lot, which could cause or contribute to the reported complaint condition.A review of other manufacturers complaint files and related trending data for similar incidents was performed to evaluate whether other similar complaints have been received and if data suggests any adverse trends may have contributed to the complaint root cause.Condition of the returned complaint related product was evaluated to confirm the reported product complaint.As part of each complaint investigation, the manufacturer performs an analysis to determine the root-cause of the reported problem.Depending upon the information provided, a true root-cause may or may not be determined.In these cases, the most-likely cause of the reported issue would be determined.Other known issues that are inherent to the business operations, typically ones that are related to shipping, handling, customer service, or other, are internally tracked and continually trended for unusual increases.As a result of the above investigation, the manufacturer has concluded the following as the root-cause or most likely root-cause of the reported complaint condition: no manufacturing defect was found.Most likely hazard experience: thread fracture during function.Most likely root cause of the failure mode / hazard: it typically results from insufficient tightening torque at placement, overloading of the abutment in function, or not retightening the abutment at prescribed intervals.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® LOCATOR® ABUTMENT 3.4MM(D) X 5MM(H)
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15392860
Report Number0001038806-2022-01404
Device Sequence Number6179285
Product Code DZE
UDI-Device Identifier00844868006350
UDI-Public(01)00844868006350(17)251120(10)1241563
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K012911
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/20/2025
Device Model NumberIMLOA005
Device Catalogue NumberIMLOA005
Device Lot Number1241563
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-