It was reported to siemens that a malfunction occurred while operating the cios alpha system.During an interventional procedure with a patient in critical condition, the user reported that the brake on the system could not be released.The clinical procedure had to be terminated.Siemens has requested additional information in order to conduct an investigation of the reported event.However, additional information was received that the involved patient had not been injured.The reported event in japan.
|
|
H10: siemens completed the investigation of the reported event.The investigation was performed based on expert discussions (considering complaint description, customer service reports, system history, c-part).During an interventional procedure, the c-arm stand brake could not be released.The procedure was continued and finished using the imperfect system, while the system was moved with the help of three people.The patient was not injured due to the event.The onsite service intervention revealed a malfunctioning brake cable rocker, which was exchanged as part of service activity.The brake cable rocker enables the wheels to lock and unlocked while the user activates/deactivates the brakes.According to the available information, the typical locking noise could no longer be heard after very little use and unlocking was prevented.This error could be simulated with the malfunctioning part in a factory test.Detailed investigation of the malfunctioning part revealed that one dimension was out of tolerance (root cause).Inventories were reviewed.No further problems were found.Therefore, this issue is considered an isolated case.The affected part was replaced as part of service activity, which resolved the problem.The spare parts consumption was checked.Any accumulation of faults or even a possible general fault that would require corrective action of the installed base could not be determined by the investigation.The incident was not classified as an adverse event after detailed investigation, because neither serious injury, death nor an unexpected or prolonged hospitalization of the patient or any other person occurred or could be expected.The corresponding probability according to the risk analysis is in the range of inconceivable.Internal id (b)(6).
|