|
SMITH & NEPHEW, INC. REF THREADED HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
| Model Number |
71336500 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during thr surgery, the ref threaded hole cover went through to back of the shell, when it was inserted to the shell.The shell was retrieved to remove the hole cover, then, the shell was placed again with a back up hole cover.The surgeon used screws to assure the shell fixation and used the same screw holes to re-insert the shell in the acetabulum.Surgery was performed, after a a non significant delay, with a smith and nephew back up device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.A dimensional inspection cannot be attempted, as the device is too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Possible probable cause could include but not limited to the user error during insertion.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown but reported as unharmed.The patient impact could not be determined based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that the correct selection of the implant is extremely important.Failure to use the optimum-sized component may result in loosening, bending, cracking, or fracture of the component and/or bone, resulting in revision surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used, size selected or user/procedural variance.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|