• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE 2000 XPI CALCITONIN (CAL); IMMULITE 2000 CALCITONIN (CAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE 2000 XPI CALCITONIN (CAL); IMMULITE 2000 CALCITONIN (CAL) Back to Search Results
Model Number N/A
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed discordant patient sample results when using immulite 2000 xpi calcitonin (cal) lot 325, in comparison to repeat results.Results for this sample were not reported to the physician.There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant immulite 2000 xpi calcitonin (cal) results.
 
Additional Manufacturer Narrative
A united states customer contacted the siemens customer care center (ccc) to report discordant immulite 2000 xpi calcitonin (cal) results when compared to repeat testing on the same instrument.It is unknown which result is correct (initial or repeat result).The limitations section of the instructions for use states the following: "for diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings." siemens continues to investigate.
 
Additional Manufacturer Narrative
Siemens filed the initial mdr 1219913-2022-00279 on 09-september-2022.Additional information 04-january-2023: siemens has reviewed a customer complaint regarding a discordant result of a patient sample with immulite 2000 calcitonin lot 325 from (b)(6) 2022.Additionally, the customer expressed concerns regarding the assay's accuracy and performance overtime.The initial result of this sample was 2.42 pmol/l, with repeat results at 1.62 pmol/l and 4.68 pmol/l.Siemens reviewed complaint database and did not find any complaints regarding the calcitonin assay from other customer sites.Siemens also reviewed kit release data of lot 325 and it met all manufacturer specifications for acceptance.Siemens then reviewed customer data from (b)(6) 2022 to (b)(6) 2022 and did not find any performance issues with the assay.The immulite 2000 calcitonin assay does not have a precision guideline or expectation for results below 3.36 pmol/l.Review of data indicates customer precision at the lower doses is within normal expectations for the calcitonin assay.There is no indication that siemens has a confirmed product issue with imprecision on the calcitonin assay.Siemens cannot rule out site specific factors as a contributing factor to the discordant result from (b)(6) 2022.Siemens has not identified a potential product issue.Customer is operational.The immulite 2000 calcitonin lot 325 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.Note: in section h6, the investigation findings and investigation conclusions codes were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMMULITE 2000 XPI CALCITONIN (CAL)
Common Device Name
IMMULITE 2000 CALCITONIN (CAL)
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
gwynedd, llanberis ll55 4el
UK 
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
uk registration #: 3002806944
gwynedd, llanberis ll55 4el
UK  
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
7818564812
MDR Report Key15393421
Report Number1219913-2022-00279
Device Sequence Number14703996
Product Code JKR
UDI-Device Identifier00630414961156
UDI-Public00630414961156
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K023304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model NumberN/A
Device Catalogue Number10381446
Device Lot Number325
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-