• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL); THYROID STIMULATING HORMONE IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL); THYROID STIMULATING HORMONE IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
A discordant, depressed thyroid stimulating hormone 3-ultra (tsh3-ul) result was obtained on a patient sample on an atellica im 1600 analyzer compared to results from two alternate methods.The discordant result was not reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely depressed tsh3-ul patient result.
 
Additional Manufacturer Narrative
A customer from outside the united states reported observation of a falsely-depressed atellica im thyroid stimulating hormone 3-ultra (tsh3-ul) result which was discordant relative to alternate-method testing.The instructions for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." siemens healthineers is investigating.
 
Additional Manufacturer Narrative
The initial mdr 1219913-2022-00280 was filed on september 9, 2022.A customer from outside of the united states reported a discordant depressed thyroid stimulating hormone 3-ultra (tsh3-ul) result was obtained on a patient sample on an atellica im 1600 analyzer compared to results from two alternate methods.Additional information - september 22, 2022.The sample cannot be sent to siemens for testing.Additional information - september 28, 2022.After reviewing the currently available information, the most likely cause of the undetectable tsh3-ul results is an extremely rare genetic variant.The limitations section of the instructions for use states: "as with any immuno-recognition measurement of a peptide, extremely rare genetic variants may exhibit varying degrees of detection.12".Siemens is continuing to investigate.
 
Additional Manufacturer Narrative
The initial mdr 1219913-2022-00280 was filed on september 9, 2022 and mdr supplemental 1 was filed on october 10, 2022.A customer from outside of the united states reported a discordant depressed thyroid stimulating hormone 3-ultra (tsh3-ul) result was obtained on a patient sample on an atellica im 1600 analyzer compared to results from two alternate methods.Additional information - october 12, 2022.After reviewing the available information, the most likely cause of the undetectable tsh3-ul results is an extremely rare genetic variant.The limitations section of the instructions for use states: "as with any immuno-recognition measurement of a peptide, extremely rare genetic variants may exhibit varying degrees of detection.12".A product problem was not identified.The customer requires no further support.In section h6, the investigation finding, and investigation conclusion codes were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL)
Common Device Name
THYROID STIMULATING HORMONE IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
333 coney street
east walpole MA 02032
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
7818564812
MDR Report Key15393440
Report Number1219913-2022-00280
Device Sequence Number5007190
Product Code JLW
UDI-Device Identifier00630414600178
UDI-Public00630414600178
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K150403
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number10995704
Device Lot Number392
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2022
Supplement Date Received by Manufacturer09/22/2022
10/12/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/10/2022
10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-