SYNTHES GMBH SPEED COMPRESSION IMPLANT KIT 09X10X10MM; STAPLE, FIXATION, BONE Back to Search Results
Model Number SE-0910
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on august 16, 2022, the staple speed compression implant kit 09x10x10mm was off the handle.This was discovered before the surgery while going through inventory.There was no patient involvement.This report involves one speed 9x10x10mm implant.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter occupation: reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: product code #: se-0910 synthes lot #: bse180746 manufacturing site: biomedical enterprises (bme) release to warehouse date: 31aug2018 expiration date: 01aug2023 supplier: (b)(4).A manufacturing record evaluation was performed for the non-sterile article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that speed compression implant kit 09x10x10mm.Was received in its original sealed package and the implant was deployed from the inserter and was freely moving inside the package, confirming the complaint condition.No other issues were identified with the returned device.A dimensional inspection for the speed compression implant kit 09x10x10mm as the complaint condition was confirmed via visual inspection and the failure mode was known to be related to historical packaging configuration.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the speed compression implant kit 09x10x10mm would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that the most probable root cause for this type of complaint is the historical packaging design, with transit/handling being a contributing factor.No manufacturing related issue was identified and/or confirmed, therefore review to the specific prm and prm line is not required.However, change order was released with an update to the packaging traveler for speed implants, indicating to package the loaded inserter with the release button facing down, instead of facing up, to avoid the tray pushing on.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEED COMPRESSION IMPLANT KIT 09X10X10MM
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15393500
Report Number8030965-2022-06593
Device Sequence Number5098069
Product Code JDR
UDI-Device Identifier00810633020111
UDI-Public00810633020111
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2023
Device Model NumberSE-0910
Device Catalogue NumberSE-0910
Device Lot NumberBSE180746
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/13/2022
10/10/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date09/13/2022
10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-