EXACTECH, INC. TRU CC FEMORAL SIZE 3 RIGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC STEM EXT 14MM X 40MM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Aseptic loosening (4883)
|
| Date of Event |
12/29/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: - logic cc femoral size 3, right (cat# 02-010-06-0330 / serial# (b)(4).- logi logic tibia ps mod insrt sz 3 13mm (cat# 02-012-35-3013 / serial# (b)(4).- lgc tibial fit tray cem sz 3f / 3t (cat# 02-012-45-3030 / serial# (b)(4).- c stem ext 14mm x 80mm (cat# 02-012-60-1480 / serial# (b)(4).- logic fit/rbk augment 1/2 block sz 3, 5mm (cat# 02-012-50-3011 / serial# (b)(4).- logic post.Aug.Block size 3, 5mm (cat# 02-010-06-0531 / serial# (b)(4).- logic fit/rbk augment 1/2 block sz 3, 5mm (cat# 02-012-50-3011 / serial# (b)(4).- cc distal fem augment sz 3, 5mm (cat# 208-05-03 / serial# (b)(4).- metaphyseal tibial cone, ml32mm (cat# 02-012-66-2000 / serial# (b)(4).- tibial stem ext.Screw (cat# 204-70-00 / serial# (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately six months post right tka, patient had a revision of the logic cc femoral construct due to loosening of the femoral components.It was revised using logic cc.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the revision reported was likely the result of both patient-related conditions and an insufficient bond between the femoral components and the bone which led to aseptic (non-infected) femoral loosening.However, this cannot be confirmed as the devices were not available for evaluation and images of the explanted devices/pre-revision radiographs were not provided.The most probable root cause associated with the reported event of ¿loosening - femoral¿ is associated with weakened integration of the femoral component at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b2, d1, d4, d8, g1, h4, h6: clinical code, component code, type of investigation, investigation conclusions.P class- the revision reported was likely the result of both patient-related conditions and an insufficient bond between the femoral components and the bone which led to aseptic (non-infected) femoral loosening.However, this cannot be confirmed as the devices were not available for evaluation and images of the explanted devices/pre-revision radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|