Model Number 3222
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The grasper was forwarded to quality engineering for evaluation.The device was visually examined, the grasper appears undamaged, the shrink tube was in a stretched condition.The distal end of shrink tube is bell-mouthed and appears to have been caught in the slot end of the jaw.It is suspected that the shrink tube was either not completely shrunk or was pulled forward during the procedure and subsequently stretched with the opening of the jaw.
 
Event or Problem Description
Heat shrink fell off during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEW FENESTRATED GRASPER TIP, DISPOSABLE
Common Device Name
MANUAL DETACHABLE SURGICAL INSTRUMENTS
Manufacturer (Section D)
MICROLINE SURGICAL
50 dunham rd suite 1500
beverly MA 01915
Manufacturer (Section G)
MICROLINE SURGICAL
50 dunham rd suite 1500
beverly MA 01915
Manufacturer Contact
scott marchand davis
50 dunham rd suite 1500
beverly, MA 01915
9789229810
MDR Report Key15393574
Report Number1223422-2022-00007
Device Sequence Number6463825
Product Code GEI
UDI-Device Identifier00811099010067
UDI-Public00811099010067
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K981389
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3222
Device Catalogue Number3222
Device Lot Number00163128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
Patient Weight129 KG
Patient EthnicityNon Hispanic
-
-