Model Number 3L92507
Medical Device Problem Codes Use of Device Problem (1670); Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2016, the patient underwent the total hip arthroplasty with the stem and the liner for the osteoarthritis of the right hip joint.The surgery was completed successfully without any surgical delay.Two years after the surgery, although the stem started to sink, the patient did not complain of pain.The surgeon was following up the patient.In 2022, the patient complained of pain, and the stem was sunk and was retro-twisting.Therefore, the revision surgery was performed on (b)(6) 2022.The stem was loosened, and the liner was not locked to the cup.The part of the liner was floating.The stem was replaced with other company¿s stem, and the liner was replaced with a new liner.According to the surgeon, the stem seemed to be slightly small.The stem was collar-less, so that the surgeon commented that he should have chosen one size larger stem for fixation.The stem began to retro-twist with sinking.Since the liner was not broken and deformed, the liner was most likely not correctly locked at the primary surgery.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This medwatch report was initially submitted on 9/22/2022.This report is being resubmitted as this report failed submission as a result of the electronic gateway.Please reference fda ticket (b)(4) for additional details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 STD SIZE 8
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15393877
Report Number1818910-2022-17560
Device Sequence Number12293104
Product Code LZO
UDI-Device Identifier10603295168737
UDI-Public10603295168737
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190344
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L92507
Device Catalogue Number3L92507
Device Lot Number5242268
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORAIL2 STD SIZE 8; PINN MAR ETO NEUT 32IDX50OD; UNKNOWN HIP ACETABULAR CUP
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-