Model Number 1023.C
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Failure of the device delayed the intubation of the patient.Based on the reported information and the event originating in the us, the criteria for reporting an adverse event have been met.
 
Event or Problem Description
Blade did not properly light.
 
Event or Problem Description
Blade did not properly light.
 
Additional Manufacturer Narrative
Failure of the device delayed the intubation of the patient.Based on the reported information and the event originating in the us, the criteria for reporting an adverse event have been met.Complaint could not be confirmed.Inventory analysis was performed for the part in the complaint.All the inventory pulled passed the inspection.Reviewed the complaint history for the 24 months preceding this event and found 8 previous complaints for light functionality.Reviewed the risk analysis file.This associated failure mode is reduced or no visualization due to a nonfunctional light source.The severity of harm for this failure mode is considered a 9, which typically triggers a carb meeting.However, this has already been brought up at carb meetings and sunmed is going though a supplier change to address the issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SALTER LABS
Common Device Name
BX20 - MAC 3 DISPOSABLE LED BLADE
Manufacturer (Section D)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer (Section G)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key15394999
Report Number3000219639-2022-00035
Device Sequence Number4905520
Product Code CCW
UDI-Device Identifier10607411966405
UDI-Public10607411966405
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1023.C
Device Catalogue Number1023.C
Device Lot NumberA0320ADSS
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-