MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CLEO 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CLEO 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7232-24
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Product samples were received for evaluation.Visual and functional testing were performed.No discrepancies were observed during visual inspection and during functional testing of the received canula, the water passed through the connector and no leaks were observed in the assembly.The reported issue could not be confirmed.The root cause cannot be established as no fault was found.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that there was a leakage at the catheter-tubing connection during the delivery of the medication.Installation sites were changed regularly.The patient had blockages because of this leakage.The lot number of the infusion set has been changed and the issue has since resolved.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO 90 INFUSION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15395060
Report Number3012307300-2022-18364
Device Sequence Number20097850
Product Code FPA
UDI-Device Identifier30610586028414
UDI-Public30610586028414
Combination Product (Y/N)N
PMA/510(K) Number
K042172
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2023
Device Model Number21-7232-24
Device Catalogue Number21-7232-24
Device Lot Number78X023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Initial Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
-
-