| Model Number |
VICM513.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that surgeon implanted a 13.2mm vicm513.2 implantable collamer lens with a -4.0 diopter, into the patient's left eye (os) on (b)(6) 2022.The lens was later removed and replaced for a shorter length lens due to the lens being too large.Reportedly."did an exchange late last week and the patient is happy.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
D4: expiration date: 30-sep-2023 should be added to the initial mdr.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the reporter indicated that surgeon implanted a 13.2mm vicm513.2 implantable collamer lens with a -5.0 diopter, into the patient's right eye (od) on (b)(6) 2022.The lens was removed and replaced for a shorter length lens due to excessive vault and significant reduction of iridocorneal angles.Reportedly.Small cb-cb length was the cause of the event.Problem was resolved.Claim#: (b)(4).
|
| |
|