H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found no damage, scuffs, pinch marks, cracks, crazing, cuts, etc.Could cause the failure mode reported; sample received does not present y-site in its configuration.Functional testing found no leakage was found in the sample.The complaint is not confirmed.Device passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4) corrected data: d3: report source.
|