• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY S CHAGAS; TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY S CHAGAS; TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN Back to Search Results
Catalog Number 06P08-50
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed false repeat reactive alinity s chagas results on multiple donor specimens which confirmed negative by esa confirmatory testing in two different sites ribc and nybc.The customer provided a summary of results across several months and lot numbers.No individual donor results were provided.The number of false repeat reactive donors for each month were: ribc: may2022 repeat reactive=2 / esa confirmatory testing negative=2, nybc: feb2022 repeat reactive=4 / esa confirmatory testing negative=4, apr2022 repeat reactive=2 / esa confirmatory testing negative=2, may2022 repeat reactive=1 / esa confirmatory testing negative=1 there was no reported impact to donor/patient management.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, field data history of alinity s chagas reagent, lot number 27175be00.The ticket search determined that there is normal complaint activity for the likely cause lot.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.Device history record review did not identify any non-conformances or deviations with the complaint lot and complaint issue.The customer data review for alinity s chagas reagent, lot number 27175be00 was performed.The analysis included initial reactive rates (irr), repeat reactive rates (rrr) and specificity (assuming zero prevalence) at rhode island blood center (ribc), across whole blood (wb) peer sites and all us customer sites using this lot.A aggregated performance of lot 27175be00 across all customer sites and wb peer sites is within product requirements and comparable to the performance of other lots of the same assay evaluated in the comparison.The reactive rates of lots 27175be00 at ribc were within product requirements and comparable to the performance at the peer sites as well as to other lots in the field.All data review indicating that the reagent performing as expected.A review of the manufacturing documentation did not identify any issues associated with the customer¿s observation.There was no issue with reagent performance identified.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the alinity s chagas reagent, lot number 27175be00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY S CHAGAS
Common Device Name
TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15396198
Report Number3002809144-2022-00323
Device Sequence Number6463904
Product Code QHM
UDI-Device Identifier00380740117221
UDI-Public00380740117221
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2022
Device Catalogue Number06P08-50
Device Lot Number27175BE00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date09/09/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY S SYSTEM, 06P16-01, (B)(6) ; ALNTY S SYSTEM, 06P16-01, (B)(6) ; ALNTY S SYSTEM, 06P16-01, (B)(6) ; ALNTY S SYSTEM, 06P16-01, (B)(6) ; ALNTY S SYSTEM, 06P16-01, (B)(6) ; ALNTY S SYSTEM, 06P16-01, (B)(6)
Patient SexUnknown
-
-